This site is under development. For the current active site, go to bcbhub.csic.es.

Orcid Publication

Detalle de: Time reconstruction and performance of the CMS electromagnetic calorimeter

Title Time reconstruction and performance of the CMS electromagnetic calorimeter
Abstract

The resolution and the linearity of time measurements made with the CMS electromagnetic calorimeter are studied with samples of data from test beam electrons, cosmic rays, and beam-produced muons. The resulting time resolution measured by lead tungstate crystals is better than 100 ps for energy deposits larger than 10 GeV. Crystal-to-crystal synchronization with a precision of 500 ps is performed using muons produced with the first LHC beams in 2008.

Apa Reference Collaboration, C. (2010). Time reconstruction and performance of the CMS electromagnetic calorimeter. Journal of Instrumentation, 5(03), T03011-T03011. 10.1088/1748-0221/5/03/T03011
Doi 10.1088/1748-0221/5/03/T03011
Publication Date 19 Mar 2010
Publication Type Article
Researcher Lara Lloret Iglesias
Ontology
Research Groups
Tools

Suggest changes

Publications are synced automatically from ORCID, so they can't be edited here directly. Spotted something wrong? Let us know below and we'll take care of it.

← Return to Home